Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the
3M™ Littmann
®
Electronic Stethoscope Model 3200
The Model 3200 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Model 3200 can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the Model 3200
as recommended below, according to the maximum output of the communications equipment.
Rated maximum output
power of transmitter, P
[W]
Separation distance according to frequency of transmitters, d [m]
150 kHz to 80 MHz
80 Mz to 800 MHz
d = 1,2 P
800 MHz to 2,5 GHz
d = 2,3 P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.