Chapter 3. Setting Up
3.1. Tabletop Mount
The reader must be placed on a rigid and flat desk or tabletop with at least 10 cm (4 inch) free
space on both sides, 20 cm (8 inch) on rear side and 50 cm (20 inch) on front side for cassette
insertion. Its space requirements are shown below.
In case that desk or table is not flat, adjust feet under the reader for leveling. The reader
weighs approximately 15 kg (33 lbs) and requires two persons to lift. Carefully place the
reader into its final position.
DANGER
Never place the reader on the floor.
Sliding the reader may result in internal damage or misalignment
of the optics.
Install in a location that is level and stable. Installation in an
unsuitable location can cause accident, or deterioration in image
quality.
WARNING
Sliding the reader may result in internal damage or misalignment
of the optics.
External vibration or shock during scanning may affect image
quality. The scanner must be placed on rigid, flat and reinforced
desk or tableto
p
.
WARNING
Unsuitable Installation Sites
Locations with excessive humidity or dust
Locations subject to high temperature
Locations subject to shaking or vibration
Locations exposed to considerable electrical or magnetic
noise, or other forms of electromagnetic energy
Locations with poor heat radiation