User's Manual

User Manual
FireCR TM-701
16
Chapter 3. Setting Up
3.1. Tabletop Mount
The scanner must be placed on a rigid and flat desk or tabletop with at least 10 cm (4 inch) free
space on both sides, 20 cm (8 inch) on rear side and 50 cm (20 inch) on front side for cassette
insertion. Its space requirements are shown below.
Figure 4. Tabletop Mount Space Requirements (Top View)
Free space of 20 cm (8
inch) on backside, at
least, to allow the power
switch, power cord and
USB cord to be reached
by hand at all times.
Free space of 50 cm (20
inch) on front side, at
least, for cassette insertion
and ejection.
Front Side Backside
WARNING
Unsuitable Installation Sites
Locations with excessive humidity or dust
Locations subject to high temperature
Locations subject to shaking or vibration
Locations exposed to considerable electrical or magnetic
noise, or other forms of electromagnetic energy
Locations with poor heat radiation