Datasheet
Signal Tolerance Resolution Notes
0-20mA <±0,1%, ±1 digit 10uA Max load 500 Ω
0-12VDC <±0,1%, ±1 digit 10mV Max current 40mA
Technical Data
Signal Tolerance Resolution Notes
0-20mA <±0,1%, ±1 digit 10uA Ri 30 Ω
0-12VDC <±0,1%, ±1 digit 10mV Ri 2 Ω
Output
Select range with the OUT switch and adjust the signal with the 10-turn potentiometer.
The selected output signal is maintained even when the display switch is turned to IN.
The outlet is short-circuit proof, has a transient protection and a 500 mAF fuse.
Input Signal
Select range with the IN switch. The input is transient protected and has a 100 mAF fuse.
Display
Input and output signals are indicated on a 3½ digit LCD with the switch marked IN, OUT.
If the battery voltage falls below the minimum for safe operation, BAT is shown in the display.
Supply
Normally from 2 x 9V batteries. PS-Test can also be connected with a special transformer
(230V 50HZ). The batteries and fuse are mounted under the back plate of the unit
for easy access.
Other data
Working temperature 0 to +50 °C
Storage temperature -10 to +60 °C
Relative humidity max 80 %
Automatic zero
Automatic polarity
Over-range indicator
Weight 400g
Size 80 x 160 x 32 mm
Accessories
Transformer 230 V 50 Hz
Test leads (4 per set)
Special ranges can be obtained on request