Datasheet
Output Tolerance Note
0-20mA <±0,1% Max load 700ohm
Technical Data
Output ”Normal”
Simulation of current signals is obtained by setting the button marked ”2-wire/Normal”
to the high position.
Output ”2-Wire”
Simulation of two-wire transmitters is obtained by depressing the button marked
”2-wire/Normal”.
Other data
Modelno: IS09E
Working temperature 0 to +50 °C
Storage temperature -10 to +60 °C
Relative humidity max 80 %
Weight 250g
Size 65 x 130 x 25 mm
Accessories
Transformer 230 V 50 Hz or 115V 60Hz
Test leads (2 per set)
3CON Electronics AB
info@3con.se
www.3con.se
Output Tolerance Note
2-wire <±0,1% Max voltage 50V
Power:
The instrument is in operation when the button marked ”PWR” is depressed.
Power is supplied by 2x9V batteries.
Average battery life is 16 hrs at 20mA. The LED marked ”BAT” indicates the voltage.
If the voltage falls below the minimum for safe operation, the LED marked ”BAT” will not
be illuminated. A special transformer (230V) can be connected for longer test periods.
Fuse:
Both fuse and batteries are easily changed by removing the back plate at the rear
of the instrument. Fuse 50mAF
Adjustments of output signal:
When the type of current signal has been selected, the correct current is obtained by pressing
the button marked ”STEP”. The current valueis then stepped to 0,4,10,12 and 20mA.
The next pressing resets the signal and it can be stepped up again.
Bräckan Norrgården
661 96 Långserud