To change the lamp lay the microscope on its
back to reach the lid on the bottom side.
Loosen screw C of the lamp socket and push it
outwards so that it is in the position shown in
Fig.1.
Loosen screw A and open the cover.
To remove the halogen lamp, use a cloth or
similar material. Do not touch the bulb with
the bare hand.
Lift out the halogen lamp and replace it with a
new one.
Close the cover and secure it with the screw.
Push the lamp socket back into the original
position and tighten screw C.
Fig. 1 Lamp socket cover: A - knurled screw, B - ventila-
tion slot, C - securing screws of lamp-holder
4.3.2 Changing the fuse
Turn off the power switch and unplug the
mains plug.
Unscrew the fuse holder on the back of the
stand base with a screwdriver.
Replace the fuse and reinsert the holder in its
socket.
5. Storage and cleaning
Keep the microscope in a clean, dry and dust
free place.
When not in use always cover the microscope
with the dust cover.
Do not expose it to temperatures below 0°C
and above 40°C and a max. relative humidity
of over 85%.
Always turn off the light immediately after use.
Use a soft linen cloth and some ethanol to
clean the microscope.
Use a soft lens tissue to clean the optics.