Owner's manual

6
Setting and selecting measurement parameters:
Press the Channel button (the menu DIS-
PLAY SIGNAL 1 will appear on the display).
Select the desired measurement parameter
with the Rate or the Store button.
Select the mode of operation with the
Date/Time button (the menu RANGE SIGNAL
1 will appear on the display).
Select the desired mode of operation with the
Rate or the Store button.
Confirm this selection with the Date/Time
button (the menu DISPLAY SIGNAL 2 will ap-
pear on the display).
Select the desired measurement parameter
with the Rate or the Store button.
Select the desired mode of operation with the
Rate or the Store button.
Confirm the selection with the Date/Time
button (a dot will appear in front of the respec-
tive measurement parameter when operating
manually).
The 3B NETlog™ device is ready to conduct meas-
urements.
8.5
Application as a handheld measuring
device with sensors
Set up the 3B NETlog™ equipment.
Connect the sensor to the relevant input and
remove the connections of the 4-mm sockets
from the same channel.
After the automatic sensor detection PROBE DE-
TECT… has been completed, the 3B NETlog
equipment is ready to conduct measurements.
8.6 Setting the sampling rate
Press the Rate button (the options list for
SAMPLE RATE will appear).
Select the desired sampling rate with the
Rate or the Store button.
Press the Date/Time button (the settings
STORE ANALOGUE INPUT 1, STORE ANALOGUE
INPUT 2 and STORE BINARY INPUTS will appear
in succession).
Select YES or NO for each setting with the
Rate or the Store button and confirm with
the Date/Time button (the subsequent set-
ting will appear after each confirmation).
8.7 Data logger
In data logger mode, the 3B NETlog™ equipment
records the data with a pre-selected sampling rate
and saves it internally. After completing a meas-
urement, the data can be transferred onto a com-
puter for evaluation.
Calling up data logger mode:
Press the Store button (STORE will appear
with START or CLEAR START dis-
played).
Starting data logger:
When the display shows START:
Start recording data with the Date/Time
button (“BUSY STOP” will appear in the dis-
play and the measurement begins).
Stopping the data logger:
When the display shows STOP:
Stop recording data by pressing the
Date/Time button ( CLEAR START will
appear in the display).
Clearing data logger:
When the display shows CLEAR:
Press the Rate button (MEM.CLEAR? YES
will appear in the display).
Confirm by pressing the Date/Time button.
Exiting data logger mode:
When the display shows CLEAR START:
Press the Channel button.
9. Use of Ethernet port
In order to connect the 3B NETlog™ unit into a
local network using its Ethernet port, carry out the
following procedure:
9.1 Assignment of an IP address by the router
Turn off the 3B NETlog™ device and unplug the
Ethernet cable.
Press down the "STORE" button and turn on
the 3B NETlog™. The display should show the
following: